3.      Water Quality

3.1       Introduction

3.2       Emergency Discharge

3.3       Construction Phase Monitoring

3.4       Post Project Water Quality Monitoring during Operation Phase

3.5       Mitigation Measures

 

 

List of Tables

 

Table 3.1      Proposed Water Quality Monitoring Stations

Table 3.2      Analytical Method to be applied to Marine Water Quality Samples

Table 3.3      Emergency Response Plan for Discharge of Untreated or Partially Treated Effluent from PPSTW

 

 

List of Figures

 

Figure 3.1         Proposed Water Quality Monitoring Stations

 

 

Appendix

 

Appendix A      Implementation Schedule of Proposed Mitigation Measures

Appendix B      Sample Data sheets for Air and Water Quality Monitoring

Appendix E      Indicative Requirements of the Post Project Water Quality Monitoring Programme

 

 

3.                              Water Quality

 

3.1                          Introduction

 

3.1.1.1              Marine water quality monitoring is recommended to be carried out during and after any emergency discharge of untreated or partially treated effluent from the upgraded PPSTW in the event of interruption of the electrical power supply or failure of treatment units.  A framework of the emergency response has been formulated in this EM&A Manual to minimize the impacts from the emergency discharge.  Effluent quality from the upgraded PPSTW will be governed by the Water Pollution Control Ordinance (WPCO) and the associated discharge licence conditions after the Project is commissioned.

 

3.1.1.2              In this section, the requirements, methodology, equipment, monitoring locations, criteria and protocols for the monitoring and audit of water quality impacts are presented.  

 

3.2                          Emergency Discharge

 

Monitoring Locations and Schedule

 

3.2.1.1              It is recommended to set up 11 monitoring stations as shown in Figure 3.1.  The stations (B1 to B6, WSD1 to WSD2 and U2) represent the sensitive receivers which could potentially be affected by the untreated or partially treated effluent from the PPSTW. Stations NM1 and NM6 are located outside the influence zone of the emergency discharge as predicted by the water quality modelling and would unlikely be affected by the Project and will therefore serve as control stations.

 

3.2.1.2              The coordinates of the proposed monitoring stations are listed in Table 3.1.

 

Table 3.1   Proposed Water Quality Monitoring Stations

Station

Description

Easting

Northing

B1

Butterfly Beach

813608.8

825772.6

B2

Castle Peak Beach

815931.5

826571.5

B3

Kadoorie Beach

816214.2

826202.8

B4

Cafeteria Old Beach

816435.4

826055.3

B5

Cafeteria New Beach

815464.5

825330.2

B6

Golden Beach

812859.2

824359.3

WSD1

Flushing Water Intake near Butterfly Beach

813215.6

825477.7

WSD2

Flushing Water Intake near LRT Terminus

815231.0

825797.2

U2

Secondary Contact Recreation Subzone at Lung Kwu Tan

809528.7

827566.9

NM6

Control Station

807599.3

820285.4

NM1

Control Station

820256.0

823213.8

 

3.2.1.3              A six-month baseline monitoring programme covering both dry and wet seasons is proposed at a frequency of once per month to establish the baseline water quality conditions at the 11 stations after commissioning of the Project.  During each monitoring event, water samples shall be collected at different tidal status (one for high tide and one for low tide).  The baseline water quality data shall be reviewed together with the routine marine water quality data collected by EPD at the North Western waters to establish the baseline water quality conditions.  The purpose of the baseline monitoring is to establish ambient conditions when the Project is operated normally.  The baseline monitoring shall be ceased in the events of any emergency discharges.

 

3.2.1.4              During each sampling occasion, measurements shall be taken at three water depths, namely, 1 m below water surface, mid-depth and 1 m above sea bed, except where the water depth is less than 6 m, in which case the mid-depth station may be omitted.  Shall the water depth be less than 3 m, only the mid-depth station will be monitored.

 

3.2.1.5              During the operation phase of the Project, in the event of emergency discharge of untreated or partially treated effluent, daily marine water monitoring shall be conducted by the Environmental Team at the 11 monitoring stations throughout the whole emergency discharge period until the baseline water quality resumes after the normal plant operation is restored.   During each monitoring event, water samples shall be collected at both high tide and low tide (i.e. twice a day).

 

3.2.1.6              The monitoring programme for such emergency discharge event shall be conducted in the first year after commissioning of the Project.  After 1-year monitoring period, a review shall be conducted by the SOR to recommend whether such monitoring shall be continued.  The review results shall be submitted to DSD prior to the submission to EPD for approval.  Termination or any amendment to the monitoring programme shall be agreed by EPD.

 

3.2.1.7              The status and locations of water sensitive receivers and the marine activities may change after issuing this Manual.  Necessary change in the monitoring locations shall be reviewed and approved by DSD and EPD, before the commencement of the monitoring.  It is recommended that the PPSTW operator should monitor the quality (i.e. E.coli level, pH value, temperature, salinity, turbidity and dissolved oxygen) and quantity of sewage effluent discharged from the upgraded PPSTW during the marine water sampling for data interpretation.

 

3.2.1.8              It is recommended that relevant government departments including DSD, EPD, AFCD, WSD and LCSD shall be informed of any emergency discharge events.  The PPSTW operator shall maintain good communications with various concerned parties.  A list of address, email address, phone and fax number of key persons in various departments responsible for action shall be made available to the PPSTW operator.  A framework of the emergency response is provided in Table 3.3.

 

Water Quality Parameter

 

3.2.1.9              The E.coli levels shall be monitored and determined by laboratory.

 

Site Records

 

3.2.1.10          Other relevant data shall also be recorded, including monitoring location / position, time, water depth, pH value, salinity, temperature, turbidity, dissolved oxygen, tidal stages, weather conditions and any special phenomena or work underway nearby which may induce water quality impact on the sensitive receivers. A sample data record sheet based on the one presented in the EM&A Guidelines for Development Projects in Hong Kong, is shown in Appendix B.  The Environmental Team Leader may modify the data record sheet for this EM&A programme, of which the format should be approved by DSD and agreed by the IEC and SOR.

 

Monitoring Equipment

 

Dissolved Oxygen and Temperature Measuring Equipment

 

3.2.1.11          The instrument shall be a portable and weatherproof DO measuring instrument complete with cable and sensor, and use a DC power source.  The equipment shall be capable of measuring:

 

·        a DO level in the range of 0 ‑ 20 mg L-1 and 0 ‑ 200% saturation; and

·        a temperature of 0 ‑ 45 degree Celsius.

 

3.2.1.12          It shall have a membrane electrode with automatic temperature compensation complete with a cable. Sufficient stocks of spare electrodes and cables shall be available for replacement where necessary. (For example, YSI model 59 meter, YSI 5739 probe, YSI 5795A submersible stirrer with reel and cable or an approved similar instrument).

 

3.2.1.13          Shall salinity compensation not be built-in to the DO equipment, in-situ salinity shall be measured to calibrate the DO equipment prior to each DO measurement.

 

Turbidity Measurement Instrument

 

3.2.1.14          Turbidity shall be measured in-situ by the nephelometric method. The instrument shall be portable and weatherproof turbidity measuring instrument using a DC power source complete with cable, sensor and comprehensive operation manuals.  It shall have a photoelectric sensor capable of measuring turbidity between 0 - 1000 NTU. The cable shall not be less than 25m in length. The meter shall be calibrated in order to establish the relationship between NTU units and the levels of suspended solids.

 

Sampler

 

3.2.1.15          A water sampler is required.  It shall comprise a transparent PVC cylinder, with a capacity of not less than 2 litres, which can be effectively sealed with latex cups at both ends.  The sampler shall have a positive latching system to keep it open and prevent premature closure until released by a messenger when the sampler is at the selected water depth (for example, Kahlsico Water Sampler or an approved similar instrument).

 

Water Depth Detector

 

3.2.1.16          A portable, battery-operated echo sounder shall be used for the determination of water depth at each designated monitoring station.  This unit can either be hand held or affixed to the bottom of the work boat, if the same vessel is to be used throughout the monitoring programme.

 

Salinity

 

3.2.1.17          A portable salinometer capable of measuring salinity in the range of 0 - 40 parts per thousand (ppt) shall be provided for measuring salinity of the water at each monitoring location.

 

pH

 

3.2.1.18          The instrument shall consist of a potentiometer, a glass electrode, a reference electrode and a temperature-compensating device. It shall be readable to 0.1pH in a range of 0 to 14. Standard buffer solutions of at least pH 7 and pH 10 shall be used for calibration of the instrument before and after use. Details of the method shall comply with APHA, 19th ed. 4500-HTB.

 

Monitoring Position Equipment

 

3.2.1.19          A hand-held or boat-fixed type digital Differential Global Positioning System (DGPS) with way point bearing indication or other equipment instrument of similar accuracy, shall be provided and used during marine water monitoring to ensure the monitoring vessel is at the correct location before taking measurements.

 

Calibration of In-Situ Instruments

 

3.2.1.20          All in-situ monitoring instruments shall be checked, calibrated and certified by a laboratory accredited under HOKLAS or any other international accreditation scheme before use and subsequently re-calibrated at three monthly intervals throughout all stages of the water quality monitoring programme.  Responses of sensors and electrodes shall be checked with certified standard solutions before each use.  Wet bulb calibration for a DO meter shall be carried out before measurement.

 

3.2.1.21          Sufficient stocks of spare parts shall be maintained for replacements when necessary.  Backup monitoring equipment shall also be made available so that monitoring can proceed uninterrupted even when some equipment is under maintenance, calibration, etc.

 

Sample Containers and Storage

 

3.2.1.22          Water samples shall be stored and preserved in suitable containers according to the Standard Methods, APHA, and packed in ice (cooled to 4°C without being frozen) and delivered to the laboratory and analysed within 24 hours after collection.  Sufficient volume of samples shall be collected to achieve the required detection limit.

 

Laboratory Measurement / Analysis

 

3.2.1.23          Analysis of E.coli levels shall be carried out in a HOKLAS or other international accredited laboratory.  The recommended analysis method is provided in Table 3.2. The analysis shall commence within 24 hours after collection of the water samples. The laboratory shall be HOKLAS accredited for analysis of E.coli in marine water.  Detailed testing methods, pre-treatment procedures, instrument use, Quality Assurance/Quality Control (QA/QC) details (such as blank, spike recovery, number of duplicate samples per batch, etc.), detection limits and accuracy shall be submitted to DSD and the IEC for agreement and EPD for approval prior to the commencement of monitoring programme. EPD may also request the laboratory to carry out analysis of known standards provided by EPD for quality assurance. Additional duplicate samples may be required by EPD for inter laboratory calibration. Remaining samples after analysis shall be kept by the laboratory for 3 months in case repeat analysis is required. If in-house or non-standard methods are proposed, details of the method verification may also be required to submit to EPD. In any circumstance, the sample testing shall have comprehensive quality assurance and quality control programmes. The laboratory shall demonstrate the programmes to EPD or his representatives when requested.

 

Table 3.2   Analytical Method to be applied to Marine Water Quality Samples

Determinant

Suggested Method

Suggested Detection Limit

E.coli

EPD HKSAR, Wat. Sci. Tech. Vol. 35, No. 11-12 pp 409-413

1 cfu per100mL

 

Emergency Response

 

General Requirement

 

3.2.1.24          The PPSTW operator shall inform DSD, EPD, WSD and LCSD in case of any emergency discharge of untreated or partially treated effluent due to power supply failure or equipment failure in the upgraded PPSTW. A list of address, email address, phone and fax number of key persons of concerned parties responsible for action should be made available to the PPSTW operator.  The PPSTW operator shall be responsible for carrying out marine water quality monitoring within 24 hours when the discharge event occurs.  The PPSTW operator shall closely liaise with relevant parties so that the DSD, EPD, WSD and LCSD can be informed promptly of any cases of emergency discharge.  The event and action plan under emergency discharge of treated or partially treated effluent are provided in Table 3.3.

 

3.2.1.25          Under emergency discharge, it is recommended that the PPSTW operator should monitor daily quality (i.e. E.coli level, pH value, temperature, salinity, turbidity and dissolved oxygen) and quantity of sewage effluent discharged from the PPSTW during the whole water quality monitoring period for data interpretation. 

 

Table 3.3          Emergency Response Plan for Discharge of Untreated or Partially Treated Effluent from PPSTW

Event

Action Plan

Failure of main treatment units due to power supply failure or equipment failure

1.      PPSTW operator to investigate the reason of plant failure and to implement appropriate remedial measures as stipulated in the contingency / operation plan for the PPSTW.

2.      If emergency discharge is required, Plant operator to inform DSD, EPD, WSD and LCSD within 1 hour after the discharge commence.

3.      EPD / LCSD to take immediate actions to prevent any bathing and water sports activities at the beaches and secondary contact recreation subzones within the North Western WCZ.

4.      PPSTW operator to instruct the Environmental Team to commence water monitoring within 24 hours after the emergency discharge event occurs.

5.      PPSTW operator to record the effluent flow and effluent quality (i.e. E.coli level, pH value, temperature, salinity, turbidity and dissolved oxygen) during the water monitoring period.

6.      DSD to ensure appropriate remedial measures as stipulated in the contingency / operation plan are implemented.

7.      Environmental Team to conduct daily marine water monitoring (as discussed in Section 3.2.1.24) until the baseline water quality levels are restored or 7 days after the emergency discharge is ceased, whichever is the shortest.

8.      The IEC shall compare the impact monitoring data with the baseline data to identify the degree of impact caused by the emergency discharge (if any) and to determine when the normal water quality conditions are restored.  The findings shall be provided to DSD, EPD and LCSD.

 

3.2.1.26          The impact monitoring data shall be compared with the baseline data and relevant water quality objectives to identify the degree of impact caused by the emergency discharge.  For each emergency discharge event, a Water Quality Monitoring Report shall be submitted to DSD, EPD and LCSD within 10 days after the laboratory testing result for the last sample collected during the monitoring period is available.  The findings of the water quality monitoring results including data presentation, statistical analysis, discussion, conclusion and recommendation shall be provided in the Water Quality Monitoring Report. 

 

3.3                          Construction Phase Monitoring

 

Construction Site Audits

 

3.3.1.1              Implementation of regular site audits is to ensure that the recommended mitigation measures are to be properly undertaken during construction phase of the Project. It can also provide an effective control of any malpractices and therefore achieve continual improvement of environmental performance on site.

 

3.3.1.2              Site audits shall include site inspections and compliance audits.

 

Site Inspections

 

3.3.1.3              Site inspections shall be carried out by the ET and shall be based on the mitigation measures for water pollution control recommended in the implementation schedule as attached in Appendix A. In the event that the recommended mitigation measures are not fully or properly implemented, deficiency shall be recorded and reported to the site management.  Suitable actions are to be carried out to:

 

·        Record the problems and investigate the causes;

·        Issue action notes to the Contractor who is responsible for the works;

·        Implement remedial and corrective actions immediately;

·        Re-inspect the site conditions upon completion of the remedial and corrective actions; and

·        Record the event and discuss with the Contractor for preventive actions.

 

Compliance Audits

 

3.3.1.4              Compliance audits are to be undertaken to ensure that a valid discharge license has been issued by EPD prior to the discharge of effluent from the construction activities of the Project site. If monitoring of the treated effluent quality from the Works Areas is required during the construction phase of the Project, the monitoring should be carried out in accordance with the WPCO license which is under the ambit of regional office (RO) of EPD. The audit results reflect whether the effluent quality is in compliance with the discharge license requirements. In case of non-compliance, suitable actions by the ET should be undertaken to:

 

·        Notify the Contractor, IEC, SOR and DSD on the non-compliance;

·        Identify the sources of pollution;

·        Check the implementation status of the recommended mitigation measures;

·        Investigate the operating conditions of the on-site treatment systems;

·        Implement corrective and remedial actions to improve the effluent quality;

·        Increase monitoring frequency until the effluent quality is in compliance with the discharge licence requirements; and

·        Record the non-compliance and propose preventive measures.

 

3.4                          Post Project Water Quality Monitoring during Operation Phase

 

3.4.1.1              A Post Project Water Quality Monitoring (PPWQM) programme will be implemented to confirm the predictions of the water quality made in the EIA report. The extent of PPWQM programme is subject to the prevailing environmental conditions at the time before commissioning of the Project. A conservative PPWQM programme is proposed, which include one year baseline monitoring before commissioning and one year impact monitoring after commissioning. The elements of the PPM programme include monitoring of effluent quality, sediment quality, marine water quality, benthic survey and ecotoxicological assessment. The PPWQM programme is enclosed as Appendix E for reference.

 

3.5                          Mitigation Measures

 

3.5.1.1              Mitigation measures for water quality control have been recommended in the EIA Report and are listed in the implementation schedule given in Appendix A.